Home > Technical Support > Application Field of the Rockwell Test…

Application Field of the Rockwell Test

Scale

Type of Indenter

Initial Test Force

Total Test Force N

Application Field

HRA

Diamond indenter

98.07 N

10kg

588.460kg

Hard Alloys, Tempered steel with carbonized surface, Hardened Steel Plates

HRD

980.7100kg

Thin steel plates, steel with tempered surface

HRC

1471150kg

Quenched steel, tempered steel, hard cast iron

HRF

Ball indenter      φ1.5875m1/16inches

588.460kg

Cast iron, aluminum-magnesium alloy, bearing alloy, annealed copper alloy, thin and soft steel plate

HRB

980.7100kg

Soft steel, aluminum alloy, copper alloy, forgeable cast iron, annealed steel

HRG

1471150kg

Phosphorus bronze, beryllium bronze, forgeable cast iron

HRH

Ball indenter                 φ3.175mm   1/8 inches

588.460kg

Aluminum, Zinc, Lead

HRE

980.7100kg

Bearing alloy, tin, hard plastic materials

HRK

1471150kg

There are 3 Rockwell Hardness Test Scales: Scale A, Scale B and Scale C.

 

Installation of the Hardness Tester

4.1 The working conditions of the hardness tester

4.1.1 Within the room temperature of 10-30℃;

4.1.2 In the room relative humidity no superior than 65%;

4.1.3 In a surrounding without any shock or corrosive medium;

 

 

4.2 The unpacking of the hardness tester

4.2.1 Cut off the packing belts on the eternal box and unscrew out the screws on the bottom edges of the wooden box and then take off the cover and take out the accessories kit.

4.2.2 Unscrew with a spanner the 2 outer hexangular bolts M10 to separate the tester from the base place (pay attention to the safety).

4.2.3 Place the tester horizontally and steadily on the working table with the horizontality no superior than 1mm/m. while open a hole in the working table (Fig.1) so as to allow the operating lever (6) to move normally. The suggested height of the table is about 500mm.


Scan the qr codeClose